4 wiring for analog output, 1 cable specifiations, 2 wiring procedure – Yokogawa Single Channel Oxygen Analyzer System ZR22/ZR402 User Manual

Page 89: 1 cable specifications

Advertising
background image

IM 11M13A01-02E

5-10

5.4

Wiring for Analog Output

This wiring is for transmitting 4 to 20 mA DC output signals to a device, e.g. recorder.
Maintain the load resistance including the wiring resistance at 550

⍀ or less.

Shielded cable

Receiver 1

Receiver 2

Converter

ϩ
Ϫ

ϩ
Ϫ

AO1(+)

AO1(-)

AO2(+)

AO2(-)

FG

F5.8E.EPS

ZR402G

Figure 5.8 Wiring for analog output

5.4.1

Cable Specifications

For this wiring, use a 2-core or a 4-core shielded cable.

5.4.2

Wiring Procedure

(1) M4 screws are used for the terminal of the converter. Each wire of the cable should

be terminated corresponding crimp-on terminals. Ensure that the cable shield is
connected to the FG terminal of the converter.

(2) Be sure to connect “+” and “-” polarities correctly.