IM 11M13A01-02E

4-8

Instrument air

Pressure regulator

Air Set

Needle

valve

flowmeter

Reference gas

Calibration gas

Check valve

~

Zero gas cylinder

ZR40H Auto Calibration unit

Zirconia Oxygen Analyzer, Detector

ZR402G Converter

Calibration gas
unit case

F1.2E.EPS

ZR22S Separate type Explosionproof

Non-hazardous area

Hazardous area

EXA ZR402G

100 to 240 V AC

Contact input

Analog output, contact output
Digital output (HART)

*2

*1

Signal

(6-core shield cable)

Heater (2-core)

*1: Shield cable;

Use shielded signal cables, and connect the shield to the FG terminal of the

converter.

*2: When a zirconia oxygen analyzer is used, 100% N

2

gas cannot be used as the zero

gas. Use approximately 1% of O

2

gas (N

2

-based).

Figure 4.7 Piping for System 2

Blowback Piping

This piping is required when the blowback function is carried out. The piping described
below provides automatic blowback operation when the “ blowback start “ command is
entered to the converter.

Solenoid valve

Pipe junction

Blow pipe

F4.8E.EPS

Pressure reduction valve

Figure 4.8 Blowback Piping

WARNING

When mounting solenoid valves (e.g. used with options like autocalibration or
blowback) in a hazardous area, then be sure to use explosionproof solenoid valves and
appropriate wiring in explosionproof conduit.