Specifications, Warning – Yokogawa Single Channel Oxygen Analyzer System ZR22/ZR402 User Manual

Page 20

Advertising
background image

IM 11M13A01-02E

2-1

2. Specifications

2.

Specifications

This chapter describes the specifications for the following:

ZR22S

Separate-type explosionproof detector (See Section 2.2.1)

ZO21R-L

Probe protector (See Section 2.2.2)

ZR22S (0.15 m)

High-temperature separate-type detector (See Section 2.3.1)

ZO21P-H

Adapter for High temperature probe (See Section 2.3.2)

ZR402G

Separate type converter (See Section 2.4)

ZA8F

Flow setting unit (See Section 2.5.1)

ZR40H

Automatic calibration unit (See Section 2.5.2)

ZO21S

Standard gas unit (See Section 2.6)

Other Equipments (See Section 2.7)

WARNING

Requirements for explosionproof use:

The ZR22S is connected to a ZR402G or AV550G*

1

that is mounted in a non-hazardous

area.

The ambient temperature is in the range -20 to +60

ЊC. The surface temperature of the

ZR22S is not over the temperature class T2 (300

ЊC)*

2

.

*1: Refer to IM 11M12D01-01E
*2: The terminal box temperature does not exceed 150

ЊC.