IM 11M13A01-02E

8-2

Setting example 2
If the setting (for a 4 mA current) is 75% O

2

, you must set the oxygen concentration for

the maximum (20 mA) point at more than 98% O

2

(75

ϫ 1.3). (Numbers after the

decimal point are rounded up.)

5

15

25

35

45

55

65

75

85

95

5

15

25

35

45

55

65

75

Minimum oxygen concentration, %O

2

(for a minimum current of 4 mA)

Maximum oxygen concentration, %O

2

(for a maximum current of 20 mA)

Ranges over which oxygen
concentrations can be set

Outside ranges

F8.0E.EPS

8.1.3

Entering Output Damping Constants

If a measured value which is adversely affected by rapid changes in the measurement-
gas oxygen concentration is used as the basis for control, frequent on-off actions of the
output may result. To avoid this, the converter allows the setting of output damping
constants ranging from 0 to 255 seconds. Select the appropriate output damping constant
from the numeric-data entry display. To set 30 seconds, enter 030.

8.1.4

Selection of Output Mode

You can select linear or logarithmic output mode. The former provides a linear relation-
ship between the analog output signal and oxygen concentration. The latter provides
logarithmic characteristic. Press the [Enter] key in the output mode display. A linear/
logarithmic selection display then appears. Select the desired mode.

CAUTION

• When you select logarithmic mode, the minimum output remains constant at 0.1% O

2

,

and the minimum oxygen concentration value remains unchanged.