2.7

Other Equipments

2.7.1

Auxiliary Ejector for High Temperature (Part No. E7046EC or E7046EN)

This ejector is used where pressure of measured gas for high temperature detector is
negative. This ejector consists of an ejector assembly, a pressure gauge and a needle
valve.

Standard Specifications

Ejector Assembly
Ejector Inlet Air Pressure: 29 to 68 kPa G
Air Consumption:

Approx. 30 to 40 l/min

Suction gas flow rate: 3 to 7 l/min
Connection to high-temperature probe adapter: R 1/2
Connection:

Rc1/4 or 1/4 NPT, SUS304 (JIS)

Tube Connection:

(

␾6/␾4 mm or 1/4 inch copper tube or stainless tube)

Pressure Gauge Assembly

Type: JIS B7505, A1.5U3/8

؋75

Material in Contact with Gas: SUS316 (JIS)
Case Material: Aluminum alloy (Paint color; black)
Scale: 0 to 100 kPa G
Bushing (G3/8

؋ Rc1/4 or 1/4NPT, SUS304 (JIS))

Needle Valve

Connection: Rc1/4 or 1/4NPT
Material: SUS316 (JIS)

Rc 1/4,

л6 / л4 TUBE joint: SUS304 (JIS)

1/4 NPT, 1/4 TUBE joint: SUS304 (JIS)

T2.8E.EPS

Part No.

E7046EC

E7046EN