4 default values, Alarms setup, Oxygen alarms – Yokogawa Single Channel Oxygen Analyzer System ZR22/ZR402 User Manual

Page 126

Advertising
background image

IM 11M13A01-02E

8-9

8. Detailed Data Setting

Parameter: Oxygen

Hysteresis: 0.1%O2

Contact delay: 3 s

Setpoints

Enter

Alarms setup

F8.5E.EPS

High High alarm: OFF

Set value: 1 0 0 . 0 % O2

High alarm: ON

Set value: 1 0 0 . 0 % O2

Low alarm: OFF

Set value: 0 . 0 % O2

Low Low alarm: OFF

Set value: 0 . 0 % O2

Enter

Oxygen alarms

F8.6E.EPS

Figure 8.5 Alarms Setup Display Figure 8.6 Oxygen Alarms Display

8.3.4

Default Values

When the analyzer is delivered, or if data are initialized, the default alarm set values are
as shown in Table 8.3.

Table 8.3 Alarm Setting Default Values

Set item

Set value

Hysteresis

0.1% O

2

Delay time

3 seconds

High-high limit alarm

Off

High-high-limit alarm setpoint

100% O

2

High-limit alarm

Off

High- and low-limit alarm setpoints

100% O

2

Low-limit alarm

Off

Low-limit alarm setpoint

0% O

2

Low-low-limit alarm

Off

Low-low-limit alarm setpoint

0% O

2

T8.3E.EPS