8 setting display item, 2 checking calibration contact output -14 – Yokogawa Single Channel Oxygen Analyzer System ZR22/ZR402 User Manual

Page 102

Advertising
background image

IM 11M12A01-03E

7-10

7.8

Setting Display Item

This section briefly describes the display item settings shown in Figure 7.11, “Basic
Panel Display.” If the humidity analyzer /HS option was specified at the time of pur-
chase, the primary value has set “Humidity.” If a mixing ratio is to be measured, change
the current primary value following the steps below.

Additionally, if the “humidity” has been selected in the Detector Type Setting in Section
7.6, earlier in this manual, the primary value is set to the humidity and the secondary
and tertiary values are current outputs 1 and 2, respectively when data initialization is
performed.

F7.11E.EPS

Tag:

0.0

%H

2

O

4.00mA -Output1
4.00mA -Output2

Tag name

Primary value

Secondary value

Tertiary value

Figure 7.11 Basic Panel Display

(1) Press the Setup key in the basic panel display to display the Execution/Setup display.

Then select Maintenance in the Execution/Setup display.

(2) Select the Display setup from the Maintenance panel display (Figure 7.12). The

Display setup display (Figure 7.13) then appears.

(3) In the above Display setup display, select the Display item. The Display item display

(Figure 7.14) then appears. From this display, select the Primary value and press the
[Enter] key to display the Display item selection display (Figure 7.15).

(4) Select the Secondary and Tertiary values in the same manner as in the steps above.
(5) Consult Table 7.2, “Display Items,” enabling the selection of display items in

individual display areas.

Enter

Display setup

Calibration setup

Blow back setup

mA-output loop check

Contact check

Maintenance

F7.12.EPS

Enter

Display item

Trend graph

Auto return time: 0min

Language: English

Display setup

F7.13.EPS

Figure 7.12 Maintenance Panel Display Figure 7.13 Display Setup