5 setting passwords, Others, Purging – Yokogawa Single Channel Oxygen Analyzer System ZR22/ZR402 User Manual

Page 135: Passwords

Advertising
background image

IM 11M12A01-03E

8-23

8. Detailed Data Setting

Enter

Others

F8.21.EPS

Clock

Averaging

Fuel setup

Purging

Unit setup

Passwords

Defaults

Enter

Purging time: 0 min

Purging

F8.22E.EPS

Figure 8.21 Other Settings Figure 8.22 Purging Display

8.6.5

Setting Passwords

The analyzer enables password settings to prevent unauthorized switching from the
Execution/Setup menu to lower leve1 menu displays. Set passwords for calibration,
blowback and maintenance use and for setup use individually.

Proceed as follows:

(1) Press the Setup key in the basic panel display to display the Execution/Setup
display.
(2) Choose Setup to display the “Commissioning” (Setup) display.
(3) Choose Others and then Passwords to display the Passwords display shown in
Figure 8.20.
(4) Choose “Calibration, Blow back and Maintenance” to set passwords for

calibration,blow back and maintenance respectively.

(5) The “text entry” display then appears. Enter up to eight alphanumeric characters
as the password.
(6) In the same manner, follow steps 1 through 5 above to set a password for setup.
(7) Record passwords to manage them appropriately.

Enter

Calibration,Blow back

and Maintenance:

Commissioning:

Passwords

F8.23E.EPS

Figure 8.23 Passwords Display

If you forget a password, select Setup in the Execution/Setup display, and enter
MOON.” By doing so, you can enter the Setup display only. Then display the Passwords
and verify the set passwords.