5. Wiring

CAUTION

•Before opening the detector cover, loosen the lock screw. If the screw is not loosened

first, the screw will damage the cover, and the terminal box will require replacement.
When opening and closing the cover, remove any sand particles or dust to avoid
gouging the thread.

• Notice when closing the cover of the detector

After screwing the cover in the detector body, secure it with the lock screw.

Lock screw

Detector cover

F5.8E.EPS

Figure 5.8

5.3.3

Connection to Converter

To connect the wiring to the converter, proceed as follows:

(1) M4 screws are used for the terminal of the converter. Each cable should be termi-

nated in the corresponding size crimp-on terminals.

(2) When a rubber insulated glass braided wire is used for wiring to the detector, use a

terminal box. For wiring between the terminal box and the converter, use basically a
cable, e.g., PVC sheathed PVC insulated cable rather than wire.

Note

The above is to prevent moisture or corrosive gas from entering the converter. Where
the ambient environment of the detector and the converter is well-maintained, it is
permissible allowed to connect the wiring from the detector directly to the converter
with protection by conduits.

WARNING

This wiring is to carry power for the heater. Be careful to wire the correct terminals, and
be careful not to ground or short circuit terminals when wiring, as otherwise the
instrument may be damaged.