IM 11M12A01-03E

5-15

5. Wiring

5.8

Wiring for Contact Input

The converter can execute specified function when receiving contact signals.

To use these contact signals, wire as follows:

Converter

Terminal box

DI-1

DI-C

DI-2

Contact input 1

Contact input 2

F5.14E.EPS

Figure 5.14 Contact Input Wiring

5.8.1

Cable Specifications

Use a 2-core or 3-core cable for this wiring. Depending on the number of input(s),
determine which cable to use.

5.8.2

Wiring Procedure

(1) M4 screws are used for the terminals of the converter. Each cable should be termi-

nated in the corresponding crimp-on terminals.

(2) The ON/OFF level of this contact input is identified by the resistance. Connect a

contact input that satisfies the specifications in Table 5.2.

Table 5.2 Identification of Contact Input ON/OFF

T5.2E.EPS

Closed

Open

Resistance

200

⍀ or less

100 k

⍀ or more