5 zo21s standard gas unit, Zo21s standard gas unit -21 – Yokogawa Single Channel Oxygen Analyzer System ZR22/ZR402 User Manual

Page 40

Advertising
background image

IM 11M12A01-03E

2-21

2. Specifications

2.5

ZO21S Standard Gas Unit

This is a handy unit to supply zero gas and span gas to the detector in a system configu-
ration based on System 1. It is used in combination with the detector only during
calibration.

Standard Specifications
Function: Portable unit for calibration gas supply consisting of span gas (air) pump, zero

gas cylinder with sealed inlet, flow rate checker and flow rate needle valve.

Sealed Zero Gas Cylinders (6 provided): E7050BA

Capacity: 1 l

Filled pressure: Approx. 686 kPa G (at 35

Њ C)

Composition: 0.95 to 1.0 vo1% O

2

+N

2

based

Power Supply: l00, 110, 115, 200, 220, 240 V AC

Ϯ 10%, 50/60 Hz

Power Consumption: Max. 5 VA

Case Material: SPCC (Cold rolled steel sheet)

Point: Epoxy resin, baked

Paint Color:

Mainframe; Munsell 2.0 GY3.1/0.5 equivalent

Cover; Munsell 2.8 GY6.4/0.9 equivalent

Piping:

␾ 6ϫ␾ 4mm flexible tube connection

Span Gas: Internal pump draws in air from atmosphere, and feeds to detector.

Weight: Approx. 3 kg

ء Non CE Mark.

Model and Codes

ZO21S

-2
-3
-4
-5
-7
-8

-J
-E

200 V AC 50/60 Hz
220 V AC 50/60 Hz
240 V AC 50/60 Hz
100 V AC 50/60 Hz
110 V AC 50/60 Hz
115 V AC 50/60 Hz

T2.6E.EPS

Model

Suffix code

Option code

Description

Standard gas unit

Power
supply

Panel

Style code

*A

Japanese version

English version

Style A

Advertising