2-16

2.4

ZA8F Flow Setting Unit and ZR40H Automatic Cali-
bration Unit

2.4.1

ZA8F Flow Setting Unit

This flow setting unit is applied to the reference gas and the calibration gas in a system
configuration (System 2).

This unit consists of a flow meter and flow control valves to control the flow of calibra-
tion gas and reference air.

• Standard Specifications

FIowmeter: Calibration gas; 0.1 to 1.0 l/min. Reference air; 0.1 to 1.0 l/min.

Construction: Dust-proof and rainproof construction

Case Material: SPCC (Cold rolled steel sheet)

Painting: Baked epoxy resin, Dark-green (Munsell 2.0 GY 3.1/0.5 or equivalent)

Pipe Connections: Rc1/4 or 1/4FNPT

Reference Air pressure: Clean air supply of measured gas pressure

plus approx. 50 kPa G

(or measured gas pressure plus approx. 150kPa G when a check valve is used)

pressure at inlet of the auto-calibration unit (Maximum 300 kPaG).

Air Consumption: Approx. 1.5 l/min

Weight: Approx. 2kg

Calibration gas (zero gas,span gas) flow : 0.7 l/min (at calibration time only)

Model and Codes

ZA8F

-J

-A

T2.5E.EPS

Model

Suffix code Option code

Description

Flow setting unit

Joint

With 1/4" NPT adapter
Style A

Rc 1/4

Style code