1 parts required for piping in system 1, 2 connection to the calibration gas inlet, 3 connection to the reference gas inlet – Yokogawa Single Channel Oxygen Analyzer System ZR22/ZR402 User Manual

Page 69: Parts required for piping in system 1 -2, Connection to the calibration gas inlet -2, Connection to the reference gas inlet -2

Advertising
background image

IM 11M12A01-03E

4-2

• It is recommended to use ZH21B dust protector to protect the probe output from dust

agitation (i.e., to prevent combustible materials from entering the probe cell) where
humidity measurements are made under dusty or combustible, such as paper dust,
environment.

4.1.1

Parts Required for Piping in System 1

Check that the parts listed in Table 4.1 are ready.

Table 4.1

Detector Piping location

Parts

Note

General-use Calibration gas inlet Stop valve

Recommended by YOKOGAWA

detector

(L9852CB or G7016XH)

nipple *

Rc1/4 or 1/4 NPT

generic parts

joint for tube connection Rc1/4 (1/4NPT) for

␾6ϫ␾4mm generic parts

soft tube

Reference gas inlet (sealed up) (when piping is required, refer to section 4.1.3)

Note : Parts with marking * are used when required.

T4.1E.EPS

4.1.2

Connection to the Calibration Gas Inlet

When carrying out calibration, connect the piping (6(O.D)

ϫ4(I.D.) mm tube) from the

standard gas unit to the calibration gas inlet of the detector. First, mount a stop valve (of
a quality specified by YOKOGAWA) or a nipple (found on the open market) as illus-
trated in Figure 4.2, and mount a joint (also found on the open market) at the stop valve
tip. (The stop valve may be mounted on the detector prior to shipping the detector is
shipped.)

Note : Mount the stop valve close to the detector.

F4.2E.EPS

Nipple

Stop valve

Tube connection
joint

Figure 4.2 Connection to the calibration gas inlet

4.1.3

Connection to the Reference Gas Inlet

• Normally, no piping is required for the reference gas inlet when the equipment uses

natural convection for reference gas (models ZR22G-

ᮀ-ᮀ-ᮀ-C). Leave the plug as it

is. If the air around the detector is polluted and the necessary oxygen concentration
(21 vol%O

2

) cannot be obtained, prepare piping the same as which described in

Section 4.2, System 2.

• When the equipment uses instrument air for the reference gas, piping is required as

described in Section 4.2, System 2 (models ZR22G-

ᮀ-ᮀ-ᮀ-E).