Yokogawa Single Channel Oxygen Analyzer System ZR22/ZR402 User Manual

Page 138

Advertising
background image

IM 11M12A01-03E

9-2

Sample gas

Comparison
water vapor

ϩ

Ϫ

H

2

O

concentration
indicator

Electrode

Water vapor

100%

Air

100%

100%

H

2

O

79%

21%

O

2

N

2

y%

x%

Sample-gas composition

Zirconia element

F9.1.EPS

Figure 9.1 Schematic Diagram of Measurement Principle

(B) For the “mixing ratio” measurement
Assuming that the mixing ratio is rkg/kg, then “r” can be calculated from the value of
H

2

O vol % as follows:

r = 0.622

ϫ x/(100 – x) …………… Equation (4)

From the above equations (1), (2) and (4), we obtain:

E = -K log y/a = -K log

Ά0.622 ϫ 21/(0.622 + r)/21·

= -K log 0.622/(0.622 + r) … …… Equation (5)
where, K = Constant
With Equation (5), we can obtain the mixing ratio rkg/kg from the electromotive force.

120

100

80

60

40

20

0

0.1

1

10

100

Cell output, mV

E = -50.74*log PX/20.6

Oxygen concentration vs. cell output

Oxygen concentration PX (%O

2

)

FAE.EPS

Advertising