2 zr40h automatic calibration unit, Zr40h automatic calibration unit -18 – Yokogawa Single Channel Oxygen Analyzer System ZR22/ZR402 User Manual

Page 37

Advertising
background image

IM 11M12A01-03E

2-18

2.4.2

ZR40H Automatic Calibration Unit

This automatic calibration unit is applied to supply specified flow of reference gas and
calibration gas during automatic calibration to the detector in a system configuration
(System 3).

• Specifications

Used when auto calibration is required for the separate type and instrument air is
provided. The solenoid valves are provided as standard.

Construction: Dust-proof and rainproof construction:

NEMA4X/IP67 solenoid valve only (excluding flowmeter)

Mounting: 2-inch pipe or wall mounting, no vibration

Materials: Body; Aluminum alloy, Piping; SUS316 (JIS), SUS304 (JIS), Flowmeter; MA

(acrylic resin), Bracket ; sus304 (JIS)

Finish: Polyurethane corrosion-resistance coating, mint green (Munse11 5.6BG3.3/2.9)

Piping Connection: Refer to Model and Suffix Codes

Power Supply: 24V DC (from ZR402G), Power consumption; Approx. 1.3W

Reference Air Pressure: Sample gas pressure plus Approx. 150 kPa (690 kPa max.),

(Pressure at inlet of auto-calibration unit)

Air Consumption: Approx. 1.5 l/min

Weight: Approx. 3.5 kg

Ambient Temperature: -20 to +55

Њ C, no condensation or freezing

Ambient Humidity: 0 to 95% RH

Storage Temperature: -30 to +65

Њ C

Model and Codes

ZR40H

-R

-T

T2.5-1E.EPS

Model

Suffix code

Option code

Description

Automatic calibration unit for ZR402G

Gas piping connection

1/4" NPT

Pipe connection (G1/2)

Pg 13.5

20 mm (M20 x 1.5)

1/2 NPT

Rc 1/4

Wiring connection

-P

-G

-M

-T

-A

Always -A

-

Advertising