3-2

3.1.2

Probe Insertion Hole

Includes those analyzers equipped with a dust protector.

When preparing the probe insertion hole, the following should be taken into consider-
ation:

CAUTION

• The outside dimension of detector may vary depending on its options. Use a pipe that

is large enough for the detector. Refer to Figure 3.1 for the dimensions.

• If the detector is mounted horizontally, the calibration gas inlet and reference gas inlet

should face downwards.
When using the detector with pressure compensation, ensure that the flange gasket
does not block the reference air outlet on the detector flange. If the flange gasket
blocks the outlet, the detector cannot conduct pressure compensation. Where neces-
sary, make a notch in the flange gasket. Confirm the external dimensions of the
detector in Section 3.7 before installation.

• The sensor (zirconia cell) at the probe tip may deteriorate due to condensation as it is

always at high temperature.

(1) Do not mount the probe with the tip higher than the probe base.
(2) If the probe length is 2.5 meters or more, the detector should be mounted vertically

(nomore than a 5

°

tilt).

(3) The detector probe should be mounted at right angles to the measurement gas

flow or the probe tip should point downstream.
Figure 3.1 illustrates an example of the probe insertion hole.

100mm

100mm

F3.1E.EPS

(vertical)

Bounds of the probe
insertion hole location

Flange matches
to the detector size

(horizontal)

(Note)
When using the detector with ZH21B dust protector the diameter of the hole
should be 80mm or larger.

*1
*Note

Type

Outside diameter of detector

Standard

50.8 mm in diameter (Note)

With dust protector

80 mm in diameter or longer(Note)

Four-hole flange Eight-hole flange

JIS flange(the detector
with dust protector)

*1

When using the detector with pressure compensation, ensure that the flange gasket does
not block the reference air outlet on the detector flange. If the flange gasket blocks the
outlet, the detector cannot conduct pressure compensation. Where necessary, make a
notch in the flange gasket. Confirm the outside dimensions of the detector in Section 3.7
before installation.

Figure 3.1 Example of forming probe insertion hole