8. Detailed Data Setting

8.

Detailed Data Setting

8.1

Current Output Setting

This section describes setting of the analog output range.

8.1.1

About Input Ranges

The minimum concentration of oxygen for the minimum current (4 mA) is 0% O

2

or 6%

to 76% O

2

. The maximum concentration of oxygen for the maximum current (20 mA)

ranges from 5% to 100% O

2

, and must be at least 1.3 times the concentration of oxygen

set for the minimum.

Setting example 1

If the setting (for a 4 mA current) is 10% O

2

, you must set the oxygen concentration

for the maximum (20 mA) point at more than 13% O

2

.

Setting example 2

If the setting (for a 4 mA current) is 75% O

2

, you must set the oxygen concentration

for the maximum (20 mA) point at more than 98% O

2

or greater, (75

ϫ1.3% O

2

).

(Numbers after the decimal point are rounded up.)

5

15

25

35

45

55

65

75

85

95

5

15

25

35

45

55

65

75

Minimum oxygen concentration, %O

2

(for a minimum current of 4 mA)

Maxim

um o

xygen concentr

ation, %O

2

(f

or a maxim

um current of 20 mA)

Ranges over which oxygen
concentrations can be set

Outside ranges

FAE.EPS

Figure A Max. and Min. Oxygen Concentration Set Ranges