3 calibration – Yokogawa Single Channel Oxygen Analyzer System ZR22/ZR402 User Manual

Page 146

Advertising
background image

IM 11M12A01-03E

9-10

9.2.3

Calibration

9.2.3.1Manual Calibration

For manual calibration, consult Section 7.11, “Calibration,” earlier in this manual.

9.2.3.2Semi-automatic Calibration

To start calibration, follow these steps:

(1) Press the Setup key in the basic panel display to display the Execution/Setup display.

Then select Calibration from the Execution/Setup display. The Calibration display
shown in Figure 9.8 appears.

(2) Select Semi-auto calibration to display the Semi-automatic calibration display shown

in Figure 9.9.

(3) Select Start calibration. The display shown in Figure 9.10 appears, and then start

calibration.

Enter

Manual calibration

Semi-auto calibration

Calibration

F9.8E.EPS

Check settings

Zero gas conc: 1 . 0 0 %

Span gas conc: 2 1 . 0 0 %
Cal time: 1 0 min 0 0 s
Go to setup to change value.
Start calibration

Cancel calibration

Enter

Semi-auto calibration

F9.9E.EPS

Figure 9.8 Calibration Display Figure 9.9 Semi-automatic Calibration Display

2 1 . 0 0%

0.5min./div

1 . 0 0%

2 1 . 0 % O2

CAL. TIME

Semi-auto calibration

F9.10E.EPS

Figure 9.10 Semi-automatic Calibration Display

To start calibration using an input contact, follow these steps:

(1) Make sure that Calibration start has been selected in the Input contacts display (see

Section 8.5, earlier in this manual).

(2) Apply an input contact to start calibration.

To stop calibration midway, follow these steps:
(1) Press the Return key. If this key is pressed midway during calibration, the calibration

will stop and the output stabilization time will be set up.

(2) Press the Return key once again to return to the basic panel display and the analyzer

will be in normal measurement.