4. Piping

4.3

Piping for System 3

Piping in System 3 is illustrated in Figure 4.9. In System 3, calibration is automated;
however, the piping is basically the same as that of System 2. Refer to Section 4.2.

Adjust secondary pressure of both the air set and the zero gas reducing valve so that
these two pressures are approximately the same. The flow rate of zero and span gases
(normally instrument air) are set by a single needle valve.

After installation and wiring, check the calibration contact output (see Sec. 7.10.2), and
adjust zero gas reducing valve and calibration gas needle valve so that zero gas flow is
within the permitted range. Next check span gas calibration contact output and adjust
air set so that span gas flow is within the permitted range.

CHECK
OUT

ZERO GAS IN

AIR IN

Instrument air Approx.
1.5 l/min.

flow
meter

flow
meter

REF
OUT

*2

*2

EV1

EV2

F2.6-3E.EPS

*2: Needle valve comes with flowmeter

It is recommended to use ZH21B dust protector to protect the probe output from dust
agitation (i.e., to prevent combustible materials from entering the probe cell) where
humidity measurements are made under dusty or combustible environment.

Instrument air

Calibration gas pressure regulator

Air Set

Needle

valve

Flowmeter

Reference gas

Calibration gas

Check valve

~

Zero gas cylinder

Model ZR40H Auto Calibration unit

Separate type Zirconia
High Temperature Humidity Analyzer
Model ZR22G Detecor

Model ZR402G Converter

Calibration gas
unit case

F4.9E.EPS

EXA ZR402G

Conduit

100 to 240 V AC

Contact input
Analog output, contact output
Digital output (HART)

Signal
(6-core shield cable)

Heater (2-core shield cable)

Figure 4.9 Piping for System 3