IM 11M12A01-03E

7-7

7. Startup

7.5

Confirmation of Converter Type Setting

This converter can be used for both the Oxygen Analyzer and the Humidity Analyzer.
Before setting the operating data, be sure to check that the desired converter model has
been set.

Note that if the converter type setting is changed, the operating data that have been set
are then initialized and the default settings remain. To set the desired operating data,
follow these steps:

(1) Press the setup key.
(2) Use the

᭢key to select Setup and press the [Enter] key.

(3) In the password display, enter the [Enter] key. If the password is to be set again,

enter the new password (for details, see Section 8.6.5, “Passwords,” later in this
manual).

(4) The Setup display shown in Figure 7.6 appears. Select “Basic setup” and press the

[Enter] key.

(5) The Basic setup display shown in Figure 7.7 then appears. Confirm the currently set

converter type. If the Humidity Analyzer option /HS was selected at the time of
purchase, the converter was set for high-temperature humidity use before shipment.

(6) If the converter type is to be changed, press the [Enter] key. The display shown in

Figure 7.8 then appears.

(7) Use the

᭢key to select the type of equipment. Then press the [Enter] key to com-

plete the converter selection.

(8) If the type of converter is changed after setting the operating data, those data are then

initialized and the default settings remain. Reset the operating data to meet the new
type of equipment.

Enter

Commissioning

mA-output setup

Alarms setup

Contact setup

Others

Basic setup

F7.6E.EPS

Enter

Basic setup

This instrument is for
Humidity
Please note:changing this
Basic setting will return ALL
other setting to Default!!

Detector is ZR22

F7.7E.EPS

Figure 7.6 Commissioning (“Setup”) Figure 7.7 Basic Setup Display

Display

This instrument is for

Please note:changing this
Basic setting will return ALL
other setting to Default!!

Detector is ZR22

Enter

Basic setup

Oxygen

Humidity

F7.8E.EPS

Figure 7.8 Basic Setup