2.3.2

Function

Display Functions:

Value Display;

Displays values of the measured Oxygen concentration,

moisture quantity, mixture ratio etc.

Graph Display;

Displays trends of measured oxygen concentration moisture
quantity, mixture ratio etc.

Data Display; Displays various useful data for maintenance, such as cell

temperature, reference junction temperature, maximum/

minimum moisture quantity and the like.

Status Messages;

Indicates an alarm or error occurrence with flashing of the corre-
sponding icon. Indicates status such as warming-up, calibrating,
or the like by symbols.

Alarm, Error Display: Displays alarms such as “Abnormal moisture quantity” or

errors such as “Abnormal cell e.m.f.” when any such status
occurs.

Calibration Functions:

Auto-Calibration;

Requires the Auto-calibration Unit. It calibrates automatically at
specified intervals.

Semi-auto Calibration; Requires the Auto-calibration Unit. Input calibration direction

on the touchpanel or contact, then it calibrates automatically
afterwards.

Manual Calibration; Calibration with opening/closing the valve of calibration gas in

operation interactively with an LCD panel.

Blowback Function:

Output through the contact at a set period and time. Auto/semi-auto
selectable.

Maintenance Functions:

Can update data settings during daily operation and checking.
Display data settings, calibration data settings, blowback data settings,

current output loop check, input/output contact check.

Setup Functions:

Initial settings suit for the plant conditions when installing the con
verter. Equipment settings, current output data settings, alarm data
settings, contact data settings, other settings.

Self-diagnosis:

This function diagnoses conditions of the converter or the probe and
indicates when any abnormal condition occurs.

Password Functions:

Enter your password to operate the analyzer excepting data display.
Individual passwords can be set for maintenance and setup.