3 installation of the detector, 4 installation of zh21b dust protector, Installation of the detector -3 – Yokogawa Single Channel Oxygen Analyzer System ZR22/ZR402 User Manual

Page 50: Installation of zh21b dust protector -3, Caution

Advertising
background image

IM 11M12A01-03E

3-3

3. Installation

3.1.3

Installation of the Detector

CAUTION

• The cell (sensor) at the tip of the detector is made of ceramic (zirconia). Do not drop

the detector, as impact will damage it.

• A gasket should be used between the flanges to prevent gas leakage. The gasket

material should be heatproof and corrosion-proof, suited to the characteristics of the
measured gas

The following should be taken into consideration when mounting the general-use
detector:

<General-use detector>
(1) Make sure that the cell mounting screws (four) at the probe tip are not loose.
(2) Where the detector is mounted horizontally, the calibration gas inlet and the refer-

ence gas inlet should face downward.

3.1.4

Installation of ZH21B Dust Protector

(1) Put the gasket that is provided by the user between the flanges and mount the dust

protector in the probe insertion hole.

(2) Make sure that the cell assembly mounting screws (four) at the probe tip are not

loose.

(3) Mount the detector so that the calibration gas inlet and the reference gas inlet face

downward

F3.2E.EPS

Unit

Ϻ

mm

Reference gas inlet

Calibration gas inlet

Figure 3.2 Installation of the dust filter