6 wiring for contact output, 1 cable specifications, 2 wiring procedure – Yokogawa Single Channel Oxygen Analyzer System ZR22/ZR402 User Manual

Page 84: Wiring for contact input -15

Advertising
background image

IM 11M12A01-03E

5-12

5.6

Wiring for Contact Output

Contact outputs 1 to 3 can be freely assigned to "low limit alarm", "high limit alarm",
etc. user selectable, but the assignment of contact output 4 is fixed ("error output"). And
the action (contact closed on error output) also cannot be changed.

When using these contact outputs, install the wiring as follows:

#1 Output

#2 Output

#3 Output

#4 Output

DO-1

DO-1

DO-2

DO-2

DO-3

DO-3

Terminal box

Annunciator or the like

F5.11E.EPS

Converter

DO-4

DO-4

Figure 5.11 Contact output wiring

5.6.1

Cable Specifications

Number of wires in cable varies depending on the number of contact used.

5.6.2

Wiring Procedure

(1) M4 screws are used for the terminals of the converter. Each wire in the cable should

be terminated in the corresponding crimp-on terminal.

(2) The capacities of the contact output relay are 30 V DC 3 A, 250V AC 3 A. Connect

a load (e.g. pilot lamp and annunciator) within these limits.