Yokogawa Single Channel Oxygen Analyzer System ZR22/ZR402 User Manual

Page 162

Advertising
background image

IM 11M12A01-03E

10-14

Table 10.5 Initialization Items and Default Values

T10.5E-1.EPS

Item

Data to be initialized

Default setting

Type of equipment

Not initialized

Detector

ZR22

Measurement gas

Wet gas

Display item

1st display item

Humidity

2nd display item

Current output 1

3rd display item

Current output 2

Tag name

Deleted

Displayed data

Trend graph

Parameter

Humidity

Sampling interval

30 seconds

Upper limit (graph)

25% O

2

Lower limit (graph)

0% O

2

Automatic return time

0 min.

Language

Not initialized

Mode

Manual

Calibration procedure Span - zero

Zero-gas concentration 1.00% O

2

Span-gas concentration 21.00% O

2

Calibration data

Calibration setting Output hold time

10 min., 00 sec.

Calibration time

10 min., 00 sec.

Interval

30 days, 00 hr.

Start date

10/01/00

Start time

00:00

Mode

No function (invalid)

Blow back Blow back setting (Output) hold time

10 min., 00 sec.

Blow back time

10 min., 00 sec.

Interval

30 days, 00 hr.

Start date

01/01/00

Start time

00:00

Parameter

Humidity

mA-output 1

Min. oxygen concentration 0% O

2

mA-output 2

Max. oxygen concentration 25% O

2

Output damping

0%

Output mode

Linear

Warm-up

4 mA

Set value

4 mA

Maintenance

Previous value held

Output hold setting Set value

4 mA

Calibration, blowback

Previous value held

Set value

4 mA

Error

Preset value held

Set value

3.4 mA

Equipment

selection

Current output

data

to be continued next page