Wiring procedure

Wiring should be preformed according to the following procedure:

1. Be sure to connect the shield line to FG terminal of the converter.
2. The outer sheath of the signal line should be stripped to a length of 50 mm or less.

The outer sheath of the power cable should be stripped to a length of 20 mm or less.

3. Signals may be affected by noise emission when the signal lines, power cable and

heater cable are located in the same conduit. When using conduit, signal lines should
be installed in a separate conduit from power and heater cables.

4. Install metal blind plug(s) in unused cable connection gland(s) of the converter.
5. Metal conduit should be grounded.
6. The following cables are used for wiring:

Table 5.1 Cable specifications

T5.1E.EPS

Terminal name of converter

Name

Need for shields

Cable type

Number of wires

CELL+, CELL-

HTR TC+, HTR TC-

CJ+, CJ-

HEATER

L, N,

AO-1+, AO-1-, AO-2+, AO-2-

DO-1, DO-2, DO-3, DO-4

AC-Z, AC-S, AC-C

DI-1, DI-2, DI-C

Converter signal

Converter heater

Power supply

Analog output

Contact output

Automatic

Calibration unit

Contact input

CVVS

CVV

CVV

CVVS

CVV

CVV

CVV

6

2

2 or 3 *

2 or 4

2 to 8

3

3

Note *: When the case is used for protective grounding, use a 2-wire cable.

CAUTION

• Select suitable cable O.D. to match the cable gland size.
• Protective grounding should be connected (Class 3) grounding (the grounding

resistance is 100

⍀ or less).

• Special cable length should be taken consideration.

For detail of the HART communication, refer to IM11M12A01-51E HART Protocol
Section 1.1.2 Communication Line Requirement..