Figure c max. and min. mixing ratio set ranges, Detailed data setting – Yokogawa Single Channel Oxygen Analyzer System ZR22/ZR402 User Manual

Page 115

Advertising
background image

IM 11M12A01-03E

8-3

8. Detailed Data Setting

“Mixing ratio” setting range

The minimum mixing ratio is set to 0 kg/kg or ranges from 0.201 to 0.625 kg/kg. The
maximum “mixing ratio” setting ranges from 0.2 to 1.0 kg/kg, and must be greater than
1.3 times plus 0.187 the mixing ratio set for the minimum.

Setting example 1

If the setting (for a 4 mA current) is 0 kg/kg, you must set the maximum (20 mA)
point at more than 0.2 kg/kg.

Setting example 2

If the setting (for a 4 mA current) is 0.201 kg/kg, you must set the maximum (20
mA) point at more than 0.449 kg/kg, (0.201

ϫ 1.3 + 0.187 kg/kg). (Numbers after

the decimal point are rounded up.)

0.4

0.45

0.5

0.55

0.6

0.65

0.7

0.75

0.8

0.85

0.9

0.95

1

0.201

0.25

0.3

0.35

0.4

0.45

0.5

0.55

0.6

0.65

Maximum mixing ratio, kg/kg at 20 mA

Minimum mixing ratio, kg/kg at 4 mA

Range over which mixing
ratio can be set

Outside ranges

FCE.EPS

Figure C Max. and Min. Mixing Ratio Set Ranges