3 operational data initialization, Handling of the zo21s standard gas unit -18, Warning – Yokogawa Single Channel Oxygen Analyzer System ZR22/ZR402 User Manual

Page 161

Advertising
background image

IM 11M12A01-03E

10-13

10. Other Functions

10.3 Operational Data Initialization

Individual set data initialization enables you to return to the default values set at the time
of delivery. There are two types of initialization: an all set-data initialization and a
function-by-function initialization. Table 10.5 lists the initialization items and default
values.

To initialize the set data, follow these steps:

(1) Press the Setup key in the Basic panel display to display the Execution/Setup display.

Then choose Setup.

(2) Select Others from the “Commissioning” (Setup) display.
(3) Choose “Defaults.” The “Defaults” display shown in Figure 10.11 appears.
(4) Select your desired item to initialize in order to display the “Defaults” Display

shown in Figure 10.12.

(5) Choose Defaults start. Initialization then starts.

Display

Calibration

Blow back

mA-outputs

Alarms

Contacts

Others

All data

Enter

Defaults

F10.11E.EPS

Enter

Defaults start

Cancel

Defaults

Display

Figure 10.11 Defaults Display Figure 10.12 Initialization Start Display

WARNING

• Do NOT turn off the power during initialization. Otherwise, initialization will not be

performed properly.