3 mounting of cable gland, Connection to the converter -7, Figure 5.3 wiring connection to the converter – Yokogawa Single Channel Oxygen Analyzer System ZR22/ZR402 User Manual

Page 76: Figure 5.4 cable gland mounting

Advertising
background image

IM 11M12A01-03E

5-4

1

FG

2

AO1

(+)

3

AO1

(-)

4

AO2

(+)

5

AO2

(-)

6

CJ
(+)

7

CJ

(-)

8

TC

(+)

9

TC

(-)

10

CELL

(+)

11

CELL

(-)

12

FG

13

DI-1

14

DI-2

15

DI-C

16

AI

(+)

17

AI
(-)

18

AC-

ZERO

19

AC-

SPAN

20

AC-

COM

21

FG

22

FG

CELL

(+)

CELL

(-)

TC

(+)

TC

(-)

CJ
(+)

CJ

(-)

23

DO-1

24

DO-1

25

DO-2

26

DO-2

27

DO-3

28

DO-3

29

DO-4

30

DO-4

31

HTR

32

HTR

33

L

34

N

35

G

36

FG

HTR

HTR

100-240V AC
50/60 Hz

F5.3E.EPS

Model ZR402G Separate type Zirconia High Temperature Humidity Analyzer,
Converter

Model ZR22G Separate type Zirconia
High Temperature Humidity Analyzer, Detector

Analog

output 2

Analog

output 1

4-20 mA DC

4-20 mA DC

Digital output

Contact

input 2

Contact input 1

Contact

output 1

Contact

output 2

Contact

output 3

Contact

output 4

Solenoid valve for span gas
for automatic calibration

Solenoid valve for zero gas
for automatic calibration

Flow setting unit

*1

*1

*1

The protective grounding for the converter should be connected either the protective ground terminal in the equipment or

the ground terminal on the case.
Standard regarding grounding: Ground to earth (Class 3 grounding), ground resistance: 100

or less.

Temperature transmitter input
4 to 20mA

Figure 5.3 Wiring connection to the converter

5.1.3

Mounting of Cable Gland

For each cable connection opening of the converter, mount a conduit that matches the
thread size or a cable gland.

EXA ZR402G

Ground terminal (M4)

Adaptor for 1/2 NPT thread

F5.4E.EPS

8-G1/2, 8-1/2NPT
or the like
(Wiring connection)

Figure 5.4 Cable gland mounting