IM 11M12A01-03E

8-8

8.2.3

Output Hold Setting

To set the output hold, follow these steps:
(1) Press the Setup key in the basic panel display to display the Execution/Setup display.

Then select Setup in the Execution/Setup display. Next, select the mA-output setup

and then the mA-output preset display as shown in Figure 8.2.

Warm up: 4mA

Preset value: 4 . 0 mA

Maintenance: Hold

Preset value: 4 . 0 mA

Cal.blowback: Hold

Preset value: 4 . 0 mA

Error: Preset

Preset value: 3 . 4 mA

Enter

mA-outputs presets

F8.2E.EPS

Warm up: 4mA

Preset value: 4 . 0 mA

Maintenance: Hold

Preset value: 4.0mA

Cal.blowback: Hold

Preset value: 4 . 0 mA

Error: Preset

Preset value: 3 . 4 mA

Enter

mA-outputs presets

Non-Hold

Hold
Preset

F8.3E.EPS

Figure 8.2 mA-output Preset Display Figure 8.3

(2) From the this display

͑Fig.8.2͒, select the desired display. Figure 8.3 shows an

example of selecting Maintenance. Select the desired output status.

(3) If a preset value is selected, set the corresponding output current. If you select a

preset value just below Maintenance, on the screen, the numeric-data entry display
will appear. Enter the current value you want. To set 10 mA, type in 100 and press
the [Enter] key to complete the setting.
The setting range is from 2.4 to 21.6 mA

8.2.4

Default Values

When the analyzer is delivered, or if data are initialized, output hold is the default as
shown in Table 8.3.

Table 8.3 Output Hold Default Values

Status

Output hold (min. and max. values)

During warm-up

4 mA

Under maintenance

Holds output at value just before maintenance started.

Under calibration or blow back

Holds output at value just before starting calibration

or "blow back."

On Error occurrence

Holds output at a preset value.

T8.3E.EPS

Preset value

4 mA

4 mA

4 mA

3.4 mA