9. Calibration

9.

Calibration

The following describes the calibration procedures for the EXA

XT

ZR Zirconia High-

temperature Humidity Analyzer.

9.1

Calibration Briefs

9.1.1

Measurement Principle of Zirconia Humidity Analyzer

A solid electrolyte such as zirconia allows the conduction of oxygen ions at high
temperatures. Therefore, when a zirconia-plated element with platinum electrodes on
both sides is heated up in contact with gases having different partial-oxygen pressures on
each side, oxygen ions flow from a high partial-oxygen pressure to a low partial-oxygen
pressure, causing a voltage. When a sample gas introduced into the zirconia-plated
element with the measurement electrode, and air (21.0 vol % O

2

) is flowed through the

reference electrode, an electromotive force (mV) is produced between the two elec-
trodes, governed by Nernst’s equation as follows:

E = - RT/nF log

e

y/a ………………………… Equation (1)

where, R = Gas constant

T = Absolute temperature
n: 4

F = Faraday’s constant

y = O

2

vol % on the zirconia element measurement electrode

a = O

2

vol % to 21.0 vol % O

2

on the zirconia element reference electrode

The humidity analyzer uses a sample gas composed of water vapor and air.

(A) For the vol % H

2

O measurement

x: Assuming that H

2

O vol % in a mixed gas is measured:

y = (100 – x)

ϫ 0.21 …………………. Equation (2)

From the above equations (1) and (2), we obtain:

E = -K log y/a = -Klog [(100 – x)

ϫ0.21] /21

= - K log (1 –0.01 x) ……………… Equation (3)
where, K = Constant
Using the above equation (3), we can calculate the water vapor in vol % from the
electromotive force.