3 auto(matic) display-revert time, 5 language selection -8, Caution – Yokogawa Single Channel Oxygen Analyzer System ZR22/ZR402 User Manual

Page 155

Advertising
background image

IM 11M12A01-03E

10-7

10. Other Functions

10.1.2.4

Default Setting

When the analyzer is delivered, or if data are initialized, the set data are by default, as
shown in Table 10.3.

Table 10.3 Default Values for Graph Setting

Item

Default Value

Parameter

Humidity

Sampling period

30 seconds

Upper limit (oxygen concentration)

25%O

2

Lower limit ((oxygen concentration)

0%O

2

Upper limit (humidity)

25% H

2

O

Lower limit (humidity)

0% H

2

O

Upper limit (mixing ratio)

0.2 kg/kg

Lower limit (mixing ratio)

0 kg/kg

T10.3E.EPS

CAUTION

If a rapid change in the measured value occurs during sampling, no sampled data are
plotted on the graph. Use the graph indication tentatively. Check the output current for
accurate data.

10.1.3

Auto(matic) Display-Revert Time

While the Execution/Setup display, or any other display that is positioned lower than the
Execution/Setup display (see Figure 7.3.1, earlier in this manual), is displayed, if there is
no key entry from the touchpanel for a certain time, the current display will automati-
cally return to the basic panel display. This action is referred to as “auto return.”

The “auto return” time setting starts from no key entry to the return to automatic return.
The “auto return” time can be set from 0 to 255 minutes. If 0 is set, there will be no
automatic return. By default, the “auto return” time is set to 0 (zero).

To set the “auto return” time, follow these steps:

(1) Select the Setup key from the basic panel display to display the Execution/Setup

display. Then select Maintenance from the Execution/Setup display.

(2) Select the Display setup from the Maintenance display.
(3) Select Auto return time. The numeric-data entry display then appears.
(4) Enter your desired automatic return time. If you want to set “one hour,” enter 060.