7. Startup

7.11 Calibration

The analyzer is calibrated in such a way that the actual zero and span gases are mea-
sured and those measured values are used to agree with the oxygen concentrations in the
respective gases.

There are three types of calibration procedures available:

(1) Manual calibration conducting zero and span calibrations, or either of these calibra-

tions in turn.

(2) Semi-automatic calibration which uses the touchpanel or a contact input signal and

conducts calibration operations based on a preset calibration time and stable time.

(3) Automatic calibration conducted at preset intervals.

Manual calibration needs the ZA8F Flow Setting Unit to allow manual supply of the
calibration gases. Semi-automatic and automatic calibrations need the ZR40H Automatic
Calibration Unit to allow automatic supply of the calibration gases. The following
sections set forth the manual calibration procedures. For details on semi-automatic and
automatic calibrations, consult Chapter 9, “Calibration,” later in this manual.

7.11.1

Calibration Setup

7.11.1.1 Mode Setting

For the mode setting, do the following:

Press the Setup key in the basic panel display to display the Execution/Setup display.
Select Maintenance in the Execution/Setup display to display the Maintenance panel
display. Then select Calibration setup to display the Calibration setup display as Figure
7.16 shows. Select Mode in this panel, and then select “Manual” for the intended
calibration herein.

Enter

Mode: Manual

Points: Both

Zero gas conc: 1 . 0 0%

Span gas conc: 2 1 . 0 0%

Timing

Calibration setup

F7.16E.EPS

Figure 7.16 Calibration Setup

7.11.1.2 Calibration Setting Procedures

Select “Points” (calibration procedure) in the Calibration setup display to display the
“Span – Zero,” “Span, Zero” selection display. In this display, select “Span – Zero.”