4 default values, Output contact -12, Caution – Yokogawa Single Channel Oxygen Analyzer System ZR22/ZR402 User Manual

Page 123: Alarms setup, Humidity alarms

Advertising
background image

IM 11M12A01-03E

8-11

8. Detailed Data Setting

CAUTION

No alarm is issued when alarm is set to “off ”(disabled) . To use the alarm functions,

be sure to set the alarm on.

Parameter: Oxygen

Hysteresis: 0.1%O2

Contact delay: 3 s

Setpoints

Enter

Alarms setup

F8.5E.EPS

Parameter:

Hysteresis: O2

Contact delay: 3 s

Setpoints

Enter

Alarms setup

Oxygen

Humidity
Mixing

F8.6.EPS

Figure 8.5 Alarms Setup Display Figure 8.6 Alarms Setup Display

Parameter: Humidity

Hysteresis: 0 . 1 %H2O

Contact delay: 3 s

Setpoints

Enter

Alarms setup

F8.7.EPS

High High alarm: OFF

Set value: 1 0 0. 0 %H2O

High alarm: OFF

Set value: 1 0 0. 0 %H2O

Low alarm: OFF

Set value: 0 . 0 %H2O

Low Low alarm: OFF

Set value: 0 . 0 %H2O

Enter

Humidity alarms

F8.8.EPS

Figure 8.7 Alarms Setup Display Figure 8.8 Humidity Alarms Display

8.3.4

Default Values

When the analyzer is delivered, or if data are initialized, the default alarm set values are
as shown in Table 8.4

Table 8.4 Alarm Setting Default Values

Set item

Oxygen concentration

Humidity (amount of

Mixing ratio

moisture content)

Set value Default Set value Default

Set value Default

setting

setting

setting

Hysteresis 0.1 to 9.9% O

2

0.1% O

2

0 to 9.9% H

2

O 0.1% H

2

O 0 to 0.1 kg/kg 0.001 kg/kg

Delay time 0 to 255

3 0 to 255 3

0 to 255 3

seconds seconds seconds seconds seconds seconds

High-high limit -- Off -- Off -- Off
alarm
High-high-limit 0 to 100% O

2

100% O

2

0 to 100% 100.0% 0 to 1 kg/kg 1 kg/kg

alarm setpoint

H

2

O

H

2

O

High-limit alarm -- Off -- Off -- Off
High- and low- 0 to 100% O

2

100% O

2

0 to 100% 100.0% 0 to 1 kg/kg 1 kg/kg

limit alarm setpoints

H

2

O H

2

O

Low-limit alarm

Off

Off

Off

Low-limit alarm 0 to 100% O

2

0% O

2

0 to 100% 0.0% 0 to 1 kg/kg 0 kg/kg

setpoint H

2

O H

2

O

Low-low-limit alarm -- Off Off Off
Low-low-limit alarm 0 to 100% O

2

0% O

2

0 to 100% H

2

O 0.0% H

2

O 0 to 1 kg/kg 0 kg/kg

setpoint

T8.4E.EPS