2 system 2, System 2 -3 – Yokogawa Single Channel Oxygen Analyzer System ZR22/ZR402 User Manual

Page 16

Advertising
background image

IM 11M12A01-03E

1-3

1. Overview

1.1.2

System 2

This system is for accurate monitoring and controlling humidity when the installation
environment is polluted with gases other than the air. Instrument air (clean and dry air
of oxygen concentration 21%) is used for the reference gas and the span gas for calibra-
tion. Zero gas is fed in from a cylinder during calibration. The gas flow is controlled by
the ZA8F flow setting unit (for manual valve operation).

F1.2E.EPS

ϳ

Model ZR402G Converter

Check valve
or Stop valve

Model ZA8F flow setting unit

EXA ZR402G

Separate type Zirconia
High Temperature Humidity Analyzer,
Model ZR22G Detector

100 to 240 V AC

Air Set

Instrument air

Calibration gas

Reference

gas

Flowmeter

Needle

valve

Calibration gas pressure regulator

Calibration gas
unit case

Zero gas cylinder

Span gas (Same as Zero gas
calibration unit)

Contact input
Analog outout, contact output
Digital output (HART)

Signal
(6-core shield cable)

Heater (2-core)

Figure1.2