8. Detailed Data Setting

Note

• Measurement range switching function by an external contact input is available for

analog output1 only.

• When making a semi-automatic calibration, be sure to set semi-automatic or auto-

matic mode using the Calibration setup display.

• When carrying out “blowback,” be sure to set “blowback” in the output contact setup.
• When the unburnt gas detection signal is sent to the contact input, the converter will

cut the power supply to the heater of the detector. As a result, the heater temperature
becomes low and Error1 or Error2 happens.

8.5.2

Setting Procedure

The following are set so that semi-automatic calibration starts when input contact open
is applied to “Input1.”

Proceed as follows:

(1) Press the “Setup” key in the Basic panel display to display the Execution/Setup

display.

(2) Select “Setup” Setup in the Execution/Setup display to display the “Commission-

ing” (Setup) display.

(3) Select “Contact setup” in the “Commissioning” (Setup) display.
(4) Select “Input contacts” in the Contact setup display. The Input contacts display

then appears, as shown in Figure 8.13.

(5) Select Input1 in that display. The Input contacts display then appears, as shown in

Figure 8.14.

(6) Select Calibration start.
(7) Select Input1 “closed.” An Open or Closed selection display then appears.
(8) Choose Open.

Input1: No function

Input1 contact is active at:

Closed

Input2: No function

Input2 contact is active at:

Closed

Enter

Input contacts

F8.13E.EPS

Input1: No function

Input1 contact is active at:

Closed

Input2: No function

Input2 contact is active at:

Closed

Enter

Input contacts

No function

Cal. gas press. low
Range change
Calibration start
Process up set
Blow back start

F8.14E.EPS

Figure 8.13 Input Contacts Display Figure 8.14 Input Contacts Display

8.5.3

Default Values

When the analyzer is delivered, or if data are initialized, contact input is disabled.