4. Piping

4.2

Piping for System 2

Piping in System 2 is illustrated in Figure 4.7.

~

EXA ZR402G

Separate type Zirconia
High Temperature Humidity Analyzer
Model ZR22G Detector

Model ZR402G Converter

Stop valve
or Check valve

Model ZA8F flow setting unit

Reference

gas

Calibration gas

Needle

valve

Flowmeter

Instrument air

Air Set

Calibration gas pressure regulator

Zero gas

cylinder

Calibration gas
unit case

F4.7E.EPS

Span gas (Same as Zero gas

Calibration unit)

100 to 240 V AC

Contact input

Analog output, contact output
Digital output (HART)

Signal
(6-core shield cable)

Heater (2-core)

Figure 4.7 Piping for System 2

Piping in System 2 is as follows:

• Place a stop valve or check valve through the nipple at the calibration gas inlet of the

detector.

• It is recommended to use ZH21B dust protector to protect the probe output from dust

agitation (i.e., to prevent combustible materials from entering the probe cell) where
humidity measurements are made under dusty or combustible environment

4.2.1

Piping Parts for System 2

Check that the parts listed in Table 4.2 are ready.

Table 4.2

Detector

Piping location

Parts

Note

General-use

Calibration gas inlet

Stop valve or check valve

Recommended by YOKOGAWA

detector

(L9852CB or G7016XH)
Provided by YOKOGAWA
(K9292DN or K9292DS)

Nipple *

Rc1/4 or 1/4 NPT

generic

parts

Zero gas cylinder

User's scope

Gas pressure regulator

Recommended by YOKOGAWA
(G7013XF or G7014XF)

Joint for tube connection

Rc1/4 or 1/4 NPT

generic

parts

Reference gas inlet

Air set

Recommended by YOKOGAWA
(K9473XH/K9473XJ or
G7004XF/ K9473XG)

Joint for tube connection

Rc1/4 or 1/4 NPT

generic

parts

Note : Parts with marking * are used when required.

T4.2E.EPS