10-1

10. Other Functions

10. Other Functions

10.1 Display

10.1.1

Detailed Display

Press the Detailed-data key on the basic panel display to display the detailed operation
data as shown in Figure 10.1.

Pressing the

᭢ or ᭡ key, you can advance the page or go back to your desired page.

• Detailed-data display

There are ten panel displays for viewing detailed data. The following briefly describe the
operational data displayed on the detailed-data display.

Span gas ratio: 0 . 0 %
Zero gas ratio 1 0 0 . 0 %
Response time: 0 s
Cell robustness: life> 1year
Cell temperature: 750

Њ

C

C.J.temperature: 47

Њ

C

Tag:

Hold

F10.1E.EPS

Figure 10.1 Detailed-data Display

10.1.1.1 Span-gas and Zero-gas Correction Ratios

These are used to check for degradation of the sensor (cell). If the correction ratio is
beyond the limits as shown in Figure 10.2, the sensor should no longer be used.

These ratios can be found by calculating the data as shown below.

21.0

p1

0.51

81.92

ez

e1

es

0

C

A

B

Zero-gas ratio = (B/A) x 100 (%) Correctable range: 100

±

30%

Span-gas ratio = (C/A) x 100 (%) Correctable range: 0

±

18%

Cell
electromotive
force, mV

Calibration curve
before correction

Previous
zero-gas data

Corrected calibration curve
(theoretical calibration curve)

Zero origin

Span origin

Span-gas
concentration

Oxygen concentration (vol% O

2

)

F10.2E.EPS

Figure 10.2