2 blowback, 1 blowback setup, Operational data initialization -13 – Yokogawa Single Channel Oxygen Analyzer System ZR22/ZR402 User Manual

Page 157: Reset -17, 2 blow back, 1 blow back setup

Advertising
background image

IM 11M12A01-03E

10-9

10. Other Functions

10.2 Blow back

10.2.1

Blow back Setup

The following sections describe the blow back setup procedures required for carrying
out blow back.

10.2.1.1

Mode

There are three blow back modes available:

(1) No function – blow back disabled.
(2) Semi-automatic – in this mode, touchpanel operations or contact input signals will

start and perform blow back operations according to a preset time and output
stabilization time.

(3) Automatic – performs blow back operations automatically according to a preset

interval.

These three modes each have the specific limitations described below:

• If “No Function” is selected

No blow back operations are executed.

• If “Semi-auto” is selected, semi-automatic blow back can be executed (but no

calibration is made even when the automatic calibration startup time is reached.)

• If “Auto” is selected, automatic blow back can be executed in either the “Auto” or

“Semi-auto” mode.

To select the desired mode, follow these steps:

(1) Select the Setup key from the basic panel display to display the Execution/Setup

display. Then select Maintenance from the Execution/Setup display.

(2) Select the Blow back setup from the Maintenance display and select Mode. The

mode selection display as shown in Figure 10.8 then appears.

Enter

No function

Semi_Auto
Auto

Mode:

Blow back setup

F10.9E.EPS

Figure 10.8 Blow back Setup Display