45 system > calibrator menu (internal 50 mhz), 45 system > calibrator menu (internal 50 mhz) -69, Boonton 4540 series rf power meter – Boonton 4540 Peak Power Meter User Manual User Manual

Page 117: Operation

Advertising
background image

Boonton 4540 Series RF Power Meter

Operation

4-69

4.5.45 System > Calibrator Menu (Internal 50 MHz)

The Calibrator Menu is used to select and control the internal 50 MHz RF calibrator. It can also select the optional, external
1 GHz accessory calibrator (Model 2530). Note that this menu does not contain any items related to sensor calibration - it is
only for controlling the calibrator for use as a signal source. For sensor calibration information, refer to Section 3.5 and
Section 4.5.2 of this manual.

Main Pg2 > System > Calibrator > Cal Output

Description:

Enable or disable the RF output of the Calibrator.


Menu Type:

Multiple Choice.


Choices:

Off: Disabled

On: Provides 50 MHz CW output

SCPI Command:

OUTPut:INTernal:SIGNal <Boolean>

Main Pg 2 > System > Calibrator > Set Level

Description:

Set the output of the selected calibrator in dBm.


Menu Type:

Numeric value.

Range:

-60.0 dBm to +20.0 dbm (Internal 50 MHz)

-50.0 dBm to +20.0 dbm (External 1 GHz)

SCPI Command:

OUTPut:INTernal:LEVel:[POWer]
<numeric_value>



Main Pg 2 > System > Calibrator > Sel Calibrator

Description:

Select the active calibrator for calibration and control.
―Int‖ selects the internal 50 MHz calibrator. ―Ext‖
selects an external 1 GHz calibrator. "Ext" cannot be
selected if an external calibrator is not connected.


Menu Type:

Toggle.

Choices:

Int, Ext


SCPI Command:

None.



Main Pg 2 > System > Calibrator > Status

Description:

This menu item is not implemented at this time.

SCPI Command:

None



Fig.4-57 Calibrator Menu

Internal 50 MHz