Boonton 4540 Series RF Power Meter

Remote Operation

5-33

MARKer:POSItion:TIMe

Description:

Set or return the time (x-axis-position) of the selected marker relative to the trigger. Note that
time markers must be positioned within the time limits of the trace window in the graph display. If
a time outside of the display limits is entered, the marker will be placed at the first or last time
position as appropriate.

Syntax:

MARKer[1|2]:POSItion:TIMe <numeric_value>

Argument:

<numeric_value> = display-time in seconds (see restrictions)

Valid Modes:

Pulse and Modulated Modes.

Restrictions:

For zero delay trigger position in the center of the display , the following relationship must be
satisfied: TrigDly - (5*time/div) < MarkerTime < TrigDly + (5*time/div) where the timebase
setting is time/div