Operation

4-50

4.5.29 Markers Menu (Modulated and Pulse Modes)

The Markers menu is used to configure measurement markers at specific points on the processed measurement waveform.
Automatic measurements may then be used to retrieve measurements at the two markers and in the interval between them.
Markers are used in Pulse Mode and Modulated Mode to perform measurements at or between two time offsets relative to the
trigger. The markers can only be placed on the visible portion of the trace as defined by the timebase and trigger delay
settings.

Main Pg2 > Markers > Marker 1 | 2

Description:

Set the time (x-axis-position) of marker 1 or 2 relative
to the trigger. Note that time markers must be
positioned within the time limits of the trace window
in the graph display. If a time outside of the display
limits is entered, the marker will be placed at the first
or last time position as appropriate.


Menu Type:

Numeric Value.


Range:

Current displayed timebase limits

SCPI Command:

MARKer[1|2]:POSItion:TIMe <numeric_value>



Main Pg2 > Markers > Delta Time

Description:

Displays the result of Marker 2 – Marker 1 in seconds.
This item is read only.

Menu Type:

Numeric value.

SCPI Command:

None.

Fig. 4-40 Markers Menu

(Modulated and Pulse)