Table 6-3. statistical automatic measurements, Boonton 4540 series rf power meter – Boonton 4540 Peak Power Meter User Manual User Manual

Page 242

Advertising
background image

Boonton 4540 Series RF Power Meter

Application Notes

6-14

Table 6-3. Statistical Automatic Measurements

In the Statistical mode the following five automatic measurements are displayed in the 4540 Series Text display for enabled
input channels. The Text column lists the abbreviated forms that appear on the display screen.

TEXT

TERM

DEFINITION


Avg

Average Power:

The unweighted avg of all power samples occurring since acquisition started.


Peak

Peak Power

The highest power sample occurring since acquisition was started.


Min

Minimum Power

The lowest power sample occurring since acquisition was started. In logarithmic
units a reading below the clip level will display as down arrows.

Pk/Avg

Pk/Avg Ratio

The ratio (in dB) of the Peak Power to the Average Power.


Dyn Rng

Dynamic Range

The ratio (in dB) of the Peak Power to the Minimum Power. Displays down

arrows if the minimum power is less than the clip level in log units.



The following cursor measurements display the user set position (independent variable) and measured values (dependent
variables) where each enabled channel's movable cursor intersects the channel's CCDF trace. The position or value
measurement text for each dependent variable is displayed in the color of its channel. The independent variable is white.
Note that the intersection of the movable cursors and the CCDF traces can be moved outside the visible display area. This
does not affect the measurements in any way.


TEXT

TERM

DEFINITION


CursPwr

Cursor Power Reference

Cursor Mode - Power Ref

The reference power level in dBr set by the user to define the

measurement point on the normalized CCDF for probability in percent.


Cursor Mode - Percent

The measured power level in dBr of the normalized CCDF at the

probability in percent specified by the user.



CursPct

Cursor Percentage

Cursor Mode - Power Ref

The measured probabilty in percent of the normalized CCDF at the

reference power level specified by the user.



Cursor Mode - Percent

The probabilty in percent set by the user to define the measurement

point on the normalized CCDF for power level in dBr.






Advertising