Boonton 4540 Series RF Power Meter

Remote Operation

5-49

READ:MARKer:AVERage?

Description:

For the specified channel and marker, return the average power or voltage at the marker. The
units are the same as the specified channel.

Syntax:

READ[1|2]:MARKer[1|2]:AVERage?

Returns:

CC, average power or voltage at marker

Where CC is the measurement condition code.

Valid Modes:

Pulse and Modulated

READ:MARKer:DELTa?

Description:

For the specified channel return the difference between MK1 and MK2. The units will be the
same as marker units.

Syntax:

READ[1|2]:MARKer:DELTa?

Returns:

CC, (MK1 – MK2)

Where CC is the measurement condition code.

Valid Modes:

All

READ:MARKer:MAXimum?

Description:

For the specified channel and marker, return the maximum power or voltage at the marker. The
units are the same as the specified channel.

Syntax:

READ[1|2]:MARKer[1|2]:MAXimum?

Returns:

CC, maximum power or voltage at marker

Where CC is the measurement condition code.

Valid Modes:

Pulse and Modulated.

READ:MARKer:MINimum?

Description:

For the specified channel and marker, return the minimum power or voltage at the marker. The
units will be the same as the specified channel.

Syntax:

READ[1|2]:MARKer[1|2]:MINimum?

Returns:

CC, minimum power or voltage at marker

Where CC is the measurement condition code.

Valid Modes:

Pulse and Modulated.