Status:operation:ntransition, Status:operation:ptransition, Status:preset – Boonton 4540 Peak Power Meter User Manual User Manual

Page 199: Status:operation:ntransition -65, Status:operation:ptransition -65, Status:preset -65, Boonton 4540 series rf power meter, Remote operation

Advertising
background image

Boonton 4540 Series RF Power Meter

Remote Operation

5-65

STATus:OPERation:NTRansition

Description:

Set or return the value of the negative transition filter bitmask for the Operation status group.
Setting a bit in the positive transition filter causes a 1 to 0 (negative) transition in the
corresponding bit of the Operation Condition register to cause a 1 to be written in the associated
bit of the Operation Event register.


Syntax:

STATus:OPERation:NTRansition <numeric_value>

Argument:

<numeric_value> = 0 to 32767

Returns:

16-bit register value (0 to 32767)

Valid Modes:

Any

STATus:OPERation:PTRansition

Description:

Set or return the value of the positive transition filter bitmask for the Operation status group.
Setting a bit in the positive transition filter causes a 0 to 1 (positive) transition in the
corresponding bit of the Operation Condition register to cause a 1 to be written in the associated
bit of the Operation Event register.


Syntax:

STATus:OPERation:PTRansition <numeric_value>

Argument:

<numeric_value> = 0 to 32767

Returns:

16-bit register value (0 to 32767)

Valid Modes:

Any

STATus:PRESet

Description:

Sets SCPI enable registers and transition filters to the default state. The Operational Enable and
Questionable Enable mask registers are both cleared so an SRQ will not be generated for these
conditions. All bits for the device and questionable calibration registers are enabled. All positive
transition filters are enabled and all negative transition filters are cleared.

Syntax:

STATus:PRESet

Argument:

None

Valid Modes:

Any

Advertising