Read:interval:minfilt, Read:interval:maximum, Read:interval:minimum – Boonton 4540 Peak Power Meter User Manual User Manual

Page 182: Read:interval:pkavg, Read:interval:minfilt? -48, Read:interval:maximum? -48, Read:interval:minimum? -48, Read:interval:pkavg? -48, Boonton 4540 series rf power meter, Remote operation

Advertising
background image

Boonton 4540 Series RF Power Meter

Remote Operation

5-48

READ:INTERval:MINFilt?

Description:

For the specified channel, return the minimum filtered power or voltage in the time interval
between marker 1 and marker 2. The units will be the same as the specified channel.

Syntax:

READ[1|2]:INTERval:MINFilt?

Returns:

CC, minimum filtered power or voltage between markers

Where CC is the measurement condition code.

Valid Modes:

Pulse and Modulated

READ:INTERval:MAXimum?

Description:

For the specified channel, return the maximum instantaneous power or voltage between marker 1
and marker 2. The units will be the same as the specified channel.

Syntax:

READ[1|2]:INTERval:AVERage?

Returns:

CC, maximum instantaneous power or voltage between markers

Where CC is the measurement condition code.

Valid Modes:

Pulse and Modulated

READ:INTERval:MINimum?

Description:

For the specified channel, return the minimum instantaneous power or voltage between marker 1
and marker 2. The units will be the same as the specified channel.

Syntax:

READ[1|2]:INTERval:MINimum?

Returns:

CC, minimum instantaneous power or voltage between markers

Where CC is the measurement condition code.

Valid Modes:

Pulse and Modulated

READ:INTERval:PKAVG?

Description:

For the specified channel, return the peak-to-average ratio of the power or voltage between marker
1 and marker 2. The units are dB for logarithmic channel units or percent for linear channel units.

Syntax:

READ[1|2]:INTERval:PKAVG?

Returns:

CC, peak-to-average ratio of power or voltage between markers

Where CC is the measurement condition code.

Valid Modes:

Pulse and Modulated

Advertising