Boonton 4540 Series RF Power Meter

General Information

1-9

Table 1-2

4540 Series Performance Specifications (continued)

(Specifications are subject to change without notice)



CALIBRATION SOURCE

Internal Calibrator

Operating Modes:

Off, On CW

Frequency:

50.025 MHz ± 0.1%

Level Range:

-60 to +20 dBm

Resolution:

0.1 dB

RF Connector:

Type N

Source VSWR:

1.05 (reflection coefficient = 0.024)

Accuracy, 0C to 20C, NIST traceable:

At 0 dBm: ±0.055 dB (1.27%)
+20 to -39 dBm: ±0.075 dB (1.74%)
-40 to -60 dBm: ±0.105 dB (2.45%)


External Calibrator

Model 2530 1 GHz Calibrator (Optional accessory. See Appendix B)

Auto-calibration:

The Calibrator is used to automatically generate linearity calibration
data for peak power sensors. It is also provides test signals.



EXTERNAL INTERFACES

Remote Control:

Complies with IEEE-488.1 and SCPI version 1993.

GPIB:

Implements AH1, SH1, T6, LE0, SR1, RL1, PP0, DC1, DT1, C0, and E1.

USB:

―USB Device‖ Type B connector - does not require any power from host

LAN:

TCP/IP Ethernet Programmable interfaces;


Multi I/O:

BNC connector, user selectable for status, trigger, alarm or voltage output

Range:

0 to 10V (Analog unipolar), -10V to +10V (Analog bipolar), 0 or 5V (Logic)

Accuracy:

±100mV typical, ±200mV max, uncalibrated

Linearity:

0.1% typical

Ext Cal / VGA Out:

HDB-15 connector. External calibrator control interface for Model 2530, or
video output for VGA compatible analog RGB video monitor.