Boonton 4540 Series RF Power Meter

Remote Operation

5-89

Table 5-1. SCPI COMMAND SUMMARY (continued)


SENSe[1|2]:CORRection:CALFactor

Set correction factor in dB. <numeric_value> = -3.00 to 3.00


SENSe[1|2]:CORRection:DCYCle

Set/return duty cycle correction factor in percent.

<numeric_value> = 0.01 to 100.0%


SENSe[1|2]:CORRection:FREQuency

Set channel frequency. <numeric_value> = 0.001e9 to

110.00e9 Hz


SENSe[1|2]:CORRection:OFFSet

Set/return sensor offset value in dB. <numeric_value> = -300 to
300


SENSe[1|2]:CORRection:TEMPcomp

Set/return temperature compensation. <Boolean> = { 0, 1, OFF,
ON }


SENSe[1|2]:FILTer:STATe

Set/return filter state. <character data> = OFF, AUTO, ON

SENSe[1|2]:FILTer:TIMe

Set or return the current length of the integration filter on the
selected channel. <numeric_value> = 0.002 to 16.000 seconds


SENSe[1|2]:IMPedance

Set/return impedance reference for voltage to power in ohms.

<numeric_value> = 10.00 to 2500.00


SENSe[1|2]:PULSe:DISTal

Set/return distal parameter of the risetime. <numeric_value> =
50.00 to 100.00


SENSe[1|2]:PULSe:ENDGT

Set/return pulse gate end position in percent of pulse time duration.
<numeric_value> = 60.00 to 100.00


SENSe[1|2]:PULSe:MESIal

Set/return mesial parameter of the risetime. <numeric_value> =
10.00 to 90.00


SENSe[1|2]:PULSe:PROXimal

Set/return proximal parameter of the risetime. <numeric_value> =
0.00 to 50.00 percent


SENSe[1|2]:PULSe:STARTGT

Set/return pulse gate start position in percent of pulse timeduration.
<numeric_value> = 0.00 to 40.00


SENSe[1|2]:PULSe:UNIT

Set/return pulse definitions units. <numeric_value> = WATTS,
VOLTS


SENSe[1|2]:SENSOR:TYPE?

Returns the sensor type for the selected channel. <character data>
= NONE, CW, VOLT, PEAK


SENSe:MBUF:SIZe

Set/return size of measurement buffer for both channels.
<numeric_value> = -1 (FIFO), 0 (disabled), 1 to 1048576 points


SENSe:MBUF:RATe

Set/return modulated mode meas buffer rate for both channels.
<numeric_value> = 1.0 to 1000.0 points per second


SENSe:MBUF:POSition?

Returns the current meas buffer write index for both channels.