5-20

DISPlay:PULSe:TSPAN

(Note: This command is supplied for compatibility with the Model 4530 Power Meter)

Description:

Set or return the horizontal time span of the display in pulse mode. Time span = 10*
Time/Division. Optional units: minutes.

Syntax:

DISPlay:PULSe:TSPAN <numeric_value>

Argument:

<numeric_value> =

10e-8 to 100 sec in a 1-2-5 sequence,

300 s,

10, 20, 50, 100, 300, 600 min

Valid Modes:

Pulse mode

Note:

There are separate time bases for the modulated mode and the pulse mode. The

arguments selected are saved and restored independently by mode.

DISPlay:SCREensaver:BRIGhtness

Description:

Set or return the display backlight brightness in percent when dimmed..

Syntax:

DISPlay:SCREensaver:BRIGhtness <numeric_value>

Argument:

<numeric_value> = 0 to 100

Valid Modes:

All

DISPlay:SCREensaver:STATe

Description:

Set or return the state of the screensaver function. When ON, the screensaver dims the LCD
backlight to a specified intensity after a specified period of keyboard inactivity. GPIB activity
does not disturb display backlighting unless DISPlay:SCREensaver commands are sent.

Syntax:

DISPlay:SCREensaver:STATe <character data>

Argument:

<Boolean> = { 0, 1, OFF, ON }

Valid Modes:

All

DISPlay:SCREensaver:TIMe

Description:

Set or return the time in minutes of no activity that dims the display backlight.

Syntax:

DISPlay:SCREensaver:TIMe <numeric_value>

Argument:

<numeric_value> = 1 to 180

Valid Modes:

All