Boonton 4540 series rf power meter, Remote operation – Boonton 4540 Peak Power Meter User Manual User Manual

Page 222

Advertising
background image

Boonton 4540 Series RF Power Meter

Remote Operation

5-88

Table 5-1. SCPI COMMAND SUMMARY (continued)


READ[1|2]:ARRay:AMEAsure:POWer?

Return new Ppeak, Pavgcyc, Pavgpulse, Ptop, Pbot, Overshoot


READ[1|2]:ARRay:AMEAsure:STATistical?

Return new Pavg, Ppeak, Pmin, PkToAvgRatio, CursorPwr,
CursorPct, Sample-Count


READ[1|2]:ARRay:AMEAsure:TIMe?

Return new Freq, Period, Width, Offtime, Dcyc, Risetime, Falltime


READ[1|2]:ARRay:CW:POWer?

Return new Pavg, Pmax, Pmin, Ppulse or Pk/Avg


READ[1|2]:ARRay:MARKer:POWer?

Return new Pavg, Pmax, Pmin, P/Avg, Pmrk1, Pmrk2,
Pmrk1/Pmrk2


READ[1|2]:CW:POWer?

Return new average reading in power units


READ[1|2]:INTERval:AVERage?

Return average power between MK1 and MK2.


READ[1|2]:INTERval:MAXFilt?

Return maximum filtered power between MK1 and MK2.


READ[1|2]:INTERval:MINFilt?

Return minimum filtered power between MK1 and MK2.


READ[1|2]:INTERval:MAXimum?

Return maximum instantaneous power between MK1 and MK2.


READ[1|2]:INTERval:MINimum?

Return minimum instantaneous power between MK1 and MK2.


READ[1|2]:INTERval:PKAVG?

Return peak to average power between MK1 and MK2.

READ[1|2]:MARKer[1|2]:AVERage?

Return new reading at the specified marker.


READ[1|2]:MARKer:DELTa?

Return difference between MK2 and MK1.


READ[1|2]:MARKer[1|2]:MAXimum?

Return maximum reading at the specified marker


READ[1|2]:MARKer[1|2]:MINimum?

Return minimum reading at the specified marker


READ[1|2]:MARKer:CURsor:PERcent?

Return percent at cursor.


READ[1|2]:MARKer:CURsor:POWer?

Return relative power at cursor.


READ[1|2]:MARKer:RATio?

Return ratio between MK2 and MK1 as a percentage.


READ[1|2]:MARKer:RDELTa?

Return difference between MK1 and MK2.


READ[1|2]:MARKer:RRATio?

Return ratio between MK1 and MK2 as a percentage.


SENSe[1|2]:AVERage

Set/return trace averaging count. <numeric_value> = 1 to 16384


SENSe[1|2]:BANDwidth

Set/return sensor video bandwidth. <character data> = LOW,
HIGH


SENSe:ARRay:CALTemp?

Returns both channels’ sensor calibration temperatures.


SENSe[1|2]:CALTemp?

Returns the selected channel’s sensor calibration temperature.

Advertising