15 system subsystem, System:autoset, System:beep[:enable – Boonton 4540 Peak Power Meter User Manual User Manual

Page 204: System:beep:immediate, System:communicate:gpib:address, 15 system subsystem -70, System:autoset -70, System:beep[:enable] -70, System:beep:immediate -70, System:communicate:gpib:address -70

Advertising
background image

Boonton 4540 Series RF Power Meter

Remote Operation

5-70

5.6.15 SYSTem Subsystem

The SYSTem group of commands is used to control system-level functions not directly related to instrument measurement
performance. SYSTem commands are used to return error codes or messages from the power meter error queue, control
hardware features (backlight and key beep), access the internal clock/calendar, and configure communication parameters for
the GPIB and LAN interfaces.

SYSTem:AUTOSET

Description:

Perform the Auto Setup function to acquire a signal and display the trace in the Pulse Mode.

Syntax:

SYSTem:AUTOSET

Argument:

None

Valid Modes:

Pulse Mode is forced if not already set.

SYSTem:BEEP[:ENABle]

Description:

Set or return the status of the audible keyboard beeper.

Syntax:

SYSTem:BEEP[:ENABle] <Boolean>

Argument:

<Boolean> = { 0, 1, OFF, ON }

Valid Modes:

Any

SYSTem:BEEP:IMMediate

Description:

Causes the system to emit an audible beep. Command only; does not respond to a query.

Syntax:

SYSTem:BEEP:IMMediate

Argument:

None

Valid Modes:

Any

SYSTem:COMMunicate:GPIB:ADDRess

Description:

Set or return the GPIB bus address.

Syntax:

SYSTem:COMMunicate:GPIB:ADDRess <numeric_value>

Argument:

<numeric_value> = 1 to 30

Valid Modes:

Any